Library / Automated Plasmid Miniprep (96-well) / Troubleshooting

Troubleshooting

The failure modes that actually bite — causes and fixes.

Common issues & fixes

SymptomLikely causeFix
Low plasmid yieldOver- or under-lysis, or bead lossKeep lysis time tight; check bead resuspension and reservoir volumes.
Poor A260/280 (protein carryover)Insufficient washingConfirm both washes ran and supernatant cleared on camera; check ethanol wash volume.
RNA contaminationRNase A omitted or inactiveAdd fresh RNase A to the resuspension buffer.
Run paused mid-protocolVision flagged a plate/tip exceptionRe-seat the flagged item as prompted; the robot resumes from that step.

When not to automate this

This protocol is currently simulated — validate yields and purity against your manual miniprep baseline before production use. For a handful of preps occasionally, a manual spin-column kit may be faster than deck setup.

Verification: this protocol is simulation-verified — confirmed to run in the simulator but not yet run on hardware by RoR. Validate before using with critical samples.